Figure 28: Export Procedure Records Progress
When the records are properly transferred to the USB memory stick, a confirmation
message will be displayed on the screen.
Figure 29: Export Procedure Records Success
The device will store up to a maximum of 1000 procedure records. Once the generator
reaches it maximum record capacity, the generator will automatically delete the oldest
record to perform another therapy session.
The procedure records are saved in both a .csv and .txt format. The serial number
plus date/time, and unique number are saved as the file name. The .txt file contains all
the user viewable information stored in the procedure record (Figure 30). The .csv file
contains details of individual treatments stored in that procedure record (Figure 31).
Specifically, each comma separated line of data contains: serial number of the Delivery
Device, treatment start date/time stamp, treatment duration.
Figure 30: Example of Exported .txt File
Figure 31: Example of Exported .csv File
Black (K) ∆E ≤5.0
More Options
From the Options menu, select More Options. The More Options screen allows the
6 options that can be selected.
Figure 32: More Options Screen
System Status
The System Status screen contains information.
• Information on the generator and Delivery Device internal device identifiers.
• Software versions.
• Ability to set date and time.
• Ability to set language.
Figure 33: System Status
Set Date and Time
The Select New Date and Time screen allows the date and time of generator to be
updated from factory defaults for time zone or daylight-saving time changes. The
generator does not automatically adjust for daylight savings time changes. Valid dates
are between 1900 and the current year.
Figure 34: Set Date and Time
Set Language
The Set Language screen allows the language to be updated from English to preloaded
language options and desired number format. These settings do not change when
powering on and off the generator.
12