Arctic Sun™ Temperature Management System Specifications
Parameter
Therapy Modes
Heater Capacity
Circulating Fluid
Reservoir Capacity
Patient Probe Type
Patient Temperature Inputs
Patient Temperature Display Range
Patient Temperature Measurement Accuracy
Responses of the PCLCS (Physiologic Closed-Loop Control
System)
Patient Temperature Control Range
Water Temperature Display Range
Water Temperature Control Range (Manual)
High Water Temperature Limit
Low Water Temperature Limit
Time to heat water from 20°C to 37°C
Sound Pressure
Mains Input
Leakage Current
Operating Relative Humidity Range
Storage Relative Humidity Range
Operating Temperature Range
Storage Temperature Range
Atmospheric Pressure Range
Dimensions
Weight
Specification
Normothermia: Control Patient, Rewarm Patient
Hypothermia:
Cool Patient, Rewarm Patient
2500 BTU/hr / 750 Watts
Sterile Water
3.5 liters
YSI 400 Series compatible
Patient Temp 1: control, monitor, alarm
Patient Temp 2: monitor, alarm
10°C to 44°C
50°F to 111.2°F
0.1°C/°F increments
±0.4°C (10°C to 32°C)
±0.2°C (32°C to 38°C)
±0.4°C (38°C to 44°C)
Includes ±0.1°C external probe
Settling Time: ~4.5 hrs
Relative Overshoot: <0.5°C
Command Overshoot: <0.5°C
Response Time: Warming (max) 33°C to 37°C: ~6 hrs
Cooling 37°C to 33°C: ~2 hrs
Steady State Deviation: 0
Tracking Error: 0
Note: All values derived from testing in simulated use.
32°C to 38.5°C
89.6°F to 101.3°F
0.1°C/°F increments
3°C to 45°C / 37.4°F to 113.0°F
0.1°C/°F increments
4°C to 42°C / 39.2°F to 107.6°F
1°C/°F increments
36°C to 42°C / 96.8°F to 107.6°F
1°C/°F increments
4°C to 25°C / 39.2°F to 77°F
1°C/°F increments
8 minutes (approximate)
Alarm Tone:
70dB to 80dB at 1 meter, repeats every 10 seconds
Alert Tone:
63dB to 71dB at 1 meter, repeats every 25 seconds
Reminder Tone:
65dB at 3 meters, 0.5 seconds on/20 seconds off
100-120VAC, 50-60Hz, 11A
220-240VAC, 50-60Hz, 5.5A
<300 μA
5% to 70% non-condensing
5% to 95% non-condensing
10°C to 27°C / 50°F to 80°F
-30°C to 50°C / -20°F to 120°F
60 kPa to 110 kPa
Height: 35 inches (89 cm)
Width:
14 inches (36 cm)
Depth:
18.5 inches (47 cm)
Empty: 43 kg / 95 lbs; Filled: 47 kg / 103 lbs
E N G L I S H
9

Publicité

loading

Produits Connexes pour Medivance Simply Advanced Arctic Sun 5000

Ce manuel est également adapté pour:

Arctic sun 5000