TEMPERATURE MANAGEMENT SYSTEM
Chapter 3 – Normothermia Settings
Normothermia Settings
Use the Normothermia Settings screen to view the current settings and modify
the settings for the following parameters. To modify any parameter setting, press
the Adjust button to the right of the parameter.
Normothermia Settings screen parameters:
Water Temperature Settings
Pre-Condition Water
Manual Control
High Water Limit
Low Water Limit
Patient Temperature Settings
High Patient Alert
Low Patient Alert
Control Strategy
Display Settings
Temperature Units
Temperature Units Adjust
Patient Temp 2
To access the Normothermia Settings screen:
1) Press Adjust on the Control Patient window.
2) Press the More button on the Control Patient Adjust window.
3) The Normothermia Settings screen will be displayed.
4) To save the new settings as the current patient therapy settings, press the
Close button. For instructions on saving the settings as the system defaults,
see Advanced Setup.
NOTE:
Verify that the Patient Temperature 1 probe is correctly positioned and
correctly connected to the system. If the patient temperature change is less than
0.15°C after the first hour of therapy, the system will generate Alert 116 – Patient
Temperature 1 Change Not Detected. If Alert 116 is not acknowledged after
5 minutes, the system will generate Alarm 117 – Patient Temperature 1 Change
Not Detected. Alarm 117 will stop therapy and an audible alarm will sound.
After Alarm 117 is acknowledged, therapy will require a restart.
Chapter 4 – Hypothermia Settings
Hypothermia Settings
Use the Hypothermia Settings screen to view the current settings and modify
the settings for the following parameters. To modify any parameter setting, press
the Adjust button to the right of the parameter.
Hypothermia Settings screen parameters:
Therapy Settings
Cooling Begins
Rewarming Begins
Water Temperature Settings
Pre-Condition Water
Manual Control
High Water Limit
Low Water Limit
Patient Temperature Settings
High Patient Alert
Low Patient Alert
Control Strategy
Display Settings
Temperature Units
Temperature Units Adjust
Patient Temp 2
To access the Hypothermia Settings screen:
1) Press Adjust on the Cool Patient window or the Rewarm Patient window.
2) Press the More button on the Cool Patient Adjust window or Rewarm
Patient Adjust window.
3) The Hypothermia Settings screen will be displayed.
4) To save the new settings as the current patient therapy settings, press the
Close button. For instructions on saving the settings as the system defaults,
see Advanced Setup.
6
NOTE:
Verify that the Patient Temperature 1 probe is correctly positioned and
correctly connected to the system. If the patient temperature change is less than
0.15°C after the first hour of therapy, the system will generate Alert 116 – Patient
Temperature 1 Change Not Detected. If Alert 116 is not acknowledged after
5 minutes, the system will generate Alarm 117 – Patient Temperature 1 Change
Not Detected. Alarm 117 will stop therapy and an audible alarm will sound. After
Alarm 117 is acknowledged, therapy will require a restart.
Chapter 5 – Advanced Setup
Use the Advanced Setup screen to view the current settings and modify the
settings for the following parameters. To modify any parameter setting, press the
Adjust button to the right of the parameter.
Location / Time Settings
Language
Number Format
Current Time
Date Format
Current Date
The following functions can be initiated from the Advanced Setup screen.
Download Patient Data: The Patient Data for the last 10 (ten) cases are
stored on the Arctic Sun™ Temperature Management System hard drive.
This data is maintained when the Arctic Sun™ Temperature Management
System is powered down, or in the event of a total loss of power.
Calibration
Total Drain
Save All Settings As Default
Upload custom file
Additionally, the following information can be viewed in the
Advanced Setup screen.
Software Versions
Last Calibration Date
Next Calibration Due
To access the Advanced Setup screen:
1) Press Advanced Setup button on the Patient Therapy Selection screen.
2) The Advanced Setup screen will be displayed.
To access the Additional Protocol Selection screen:
Refer to the Arctic Sun™ Temperature Management System Help screens for
information regarding additional protocol setup.
Chapter 6 – Alarms and Alerts
The Arctic Sun™ Temperature Management System safety system continually
monitors the state of the device and the patient, and issues alarms or alerts to
notify the user of conditions that may interfere with patient safety or system
performance.
There are two types of conditions: Alarms and Alerts.
An Alarm notifies the user that a condition that may potentially pose an unsafe
situation with respect to the patient or the device. An Alarm is a High Priority
condition that requires immediate operator response.
An Alert informs the user about patient and device status without interrupting
the procedure. An Alert is a Medium Priority condition that requires prompt
operator response.
Alarms
An Alarm is denoted by an audio signal that repeats every 10 seconds until the
Alarm is cleared. The Alarm screen will appear that displays the alarm number,
alarm title, a description of the problem or conditions that triggered the alarm,
and solutions and instructions for troubleshooting and resolving the alarm
condition. If certain Alarm conditions are not acknowledged by the operator
within 2 minutes, a Reminder tone will sound. All Alarm settings are maintained
in the event of a power interruption.