Installing the optional line management clip

WARNING
Do not allow the optional line management clip to interfere with a mechanical or electronic mechanism of the product.
CAUTION
Always make sure that the clip is stable when installed.
The optional line management clip holds tubes that are hung around the product. The clip can hold four 1/2" tubes.
To install the optional line management clip:
1.
Open the clip (A)
(Figure 1-123 on page
2.
Secure the line management clip to the headboard, footboard, or siderail.
To insert a tube into the line management clip:
1. Raise the clip prong (B)
(Figure 1-123 on page
2. Insert the tube into the clip opening.
3. Release the clip prong to secure the tube into the line management clip.
WARNING
Do not pinch tubes inside the clip.
Do not clean the clip with a liquid solution.
CAUTION
Do not inset tubes that are larger than 0.75 in. into the line management clip.
Always sterilize the clip after each use.
Always make sure that the clip is stable when installed.
www.stryker.com
Accessories
1-107).
1-107).
A
Figure 1-123: Line management clip
2141-800-1 07 REV A
B
English
EN
1-107