WARNING (CONTINUED)
Do not exceed the safe working load of the Isolibrium support surface. Excess weight could cause unpredictable
safety and performance of this system.
Always use extra caution when reading radiology images taken of a patient on a support surface because internal
components can cause artifacts and distort readings.
Do not extubate or intubate patients during Lateral Rotation or Turn Assist. The functions could interfere with the
performance of the ancillary devices.
Do not zero the bed scales or weigh the patient with Lateral Rotation or Turn Assist active. Motion from the support
surface functions may adversely affect the scale system performance.
Do not arm bed exit with Lateral Rotation or Turn Assist active. The patient motion and position that results from the
support surface may adversely affect bed exit system performance.
Do not leave the patient unattended during Turn Assist.
Always deflate the Isolibrium support surface before beginning CPR.
Always lock the control panel when you leave the patient unattended. Always lock the control panel when the
patient's condition requires extra safety measures.
Do not use iBed
®
Awareness as a lock indicator for siderails. iBed
of the siderails. It is not intended to replace patient monitoring protocol.
®
The iBed
Awareness LED light bars are only intended to monitor the product status and parameter conditions. It is
not intended to replace patient monitoring protocol.
You must physically verify that the siderails are locked before arming iBed
Always securely set the footboard connector on the optional bed extender into the footboard connector slot at the
foot end of the product.
Do not pinch the power cord or cables when installing the optional bed extender.
Do not sit on the optional bed extender. This may cause the product to tip.
Do not allow the optional line management clip to interfere with a mechanical or electronic mechanism of the
product.
Do not pinch tubes inside the clip.
Do not clean the clip with a liquid solution.
Always adjust the scale or bed exit system if an option is added while the scale or bed exit system is armed.
Do not place objects that exceed 40 lb (18 kg) onto the optional monitor tray.
Do not exceed the 150 lb (68 kg) load capacity for the tray support pole.
Do not clean, service, or perform maintenance while the product is in use.
Always unplug the power cord and turn the battery switch to the OFF (O) position before cleaning, servicing, or
performing maintenance.
Always immediately unplug the power cord from the wall outlet when large spills occur near the circuit boards,
cables, and motors. Remove the patient from the product, clean up the fluid, and have service personnel completely
inspect the product. Fluids can cause unpredictable operation and decreased functionality of any electrical product.
Do not return the product to service until it is completely dry and has been thoroughly tested for safe operation.
CAUTION
Always plug the product into a wall outlet (regulated AC power source) when not in use to maintain a sufficient
battery charge and to maximize product performance while operating on battery power.
Always immediately replace batteries that have corrosion at the terminals, display cracking, have expanded or
bulging sides, or no longer can maintain a full charge.
www.stryker.com
Summary of safety precautions
2141-800-1 07 REV A
®
Awareness is only intended to detect the position
®
Awareness.