Battery modes

InTouch is equipped with two batteries that charge when the product is plugged in to a wall outlet. The battery modes
begin when the product is unplugged, or during a power failure. The InTouch battery back-up functionality activates when
you unplug the product.
WARNING
Always wash your hands after handling a battery. Battery posts, terminals and related accessories contain lead and
lead compounds, chemicals known to the State of California to cause cancer and birth defects or other reproductive
harm. Properly dispose of batteries when required.
Power save mode activates after one hour on battery power with no motion release switch activation. Bed exit, scale,
and product motion stops operating when the product enters the power save mode.
CAUTION
Always plug the product into a wall outlet (regulated AC power source) when not in use to maintain a sufficient
battery charge and to maximize product performance while operating on battery power.
Always immediately replace batteries that have corrosion at the terminals, display cracking, have expanded or
bulging sides, or no longer can maintain a full charge.
Always use only Stryker authorized batteries when replacing the batteries. Use of non-Stryker batteries may lead to
unpredictable system performance.
Upon a Battery Low alarm (Battery Low LED on Footboard and audible beep), stop using the Zoom motorized drive
and recharge the batteries immediately. Ignoring the Battery Low alarms may cause your batteries to degrade
quicker than normal and may decrease battery life.
Note: The settings for lockout controls, scale calibration data, and bed exit are preserved when the product is unplugged,
or during a power failure.
The InTouch batteries cycle through various modes of operation:
Mode
Sleep
Alarm mode before dead
Dead
Power up
Normal
www.stryker.com

Operation

Operation
Occurs if no power is being supplied to the product, if no control panel or touch screen
activity has been detected within five minutes, or if no product activity is detected while
on DC power within one hour
Occurs if the batteries are weak, bed exit is armed during sleep mode, or if no control
panel activity is detected within four hours
Occurs if power is not restored to the product or no control panel activity has been
detected within 15 minutes after the product has been in alarm mode before dead
mode, if the batteries are weak, bed exit is not armed during sleep mode, or if no
control panel activity is detected within one hour
Occurs if power is restored to the product. The product exits dead mode and enters
power up mode if the batteries are charging. After the batteries are fully charged, the
product returns to normal mode.
Occurs if power is restored to the product, control panel activity is detected, or the
touch screen is not in calibration mode