Activating and resetting CPR

EN
WARNING
Always deflate the Isolibrium support surface before beginning CPR.
To activate CPR:
1. Options to activating CPR include:
a. Pull either the patient left
end of the support surface until it locks.
b. Depress the InTouch CPR release pedal, located toward the head end of the bed to deflate the support surface
and flatten the InTouch bed frame.
c. Both of the CPR straps and the InTouch CPR release pedal can be activated
2. After you activate the CPR straps or CPR release pedal the screen displays
3. Tap OK to confirm. The screen displays CPR Activated until you reset CPR
71 on page
1-72).
Figure 1-70: CPR activated, patient left
To reset the CPR straps, pull down on the CPR straps until they snap back into their fully seated position.
To reset the InTouch CPR release pedal, tap OK
1-72
Operation
(Figure 1-70 on page
1-72), patient right or both CPR straps that are located at the foot
Figure 1-69: CPR Activated — Surface is deflating
(Figure 1-71 on page
2141-800-1 07 REV A
(Figure 1-71 on page
(Figure 1-69 on page
(Figure 1-70 on page 1-72
Figure 1-71: CPR activated, both straps and pedal
1-72).
1-72).
1-72).
and
Figure 1-
www.stryker.com