WARNING
Do not clean, service, or perform maintenance while the product is in use.
Always unplug the power cord and turn the battery switch to the OFF (O) position before cleaning, servicing, or
performing maintenance.
Always immediately unplug the power cord from the wall outlet when large spills occur near the circuit boards,
cables, and motors. Remove the patient from the product, clean up the fluid, and have service personnel completely
inspect the product. Fluids can cause unpredictable operation and decreased functionality of any electrical product.
Do not return the product to service until it is completely dry and has been thoroughly tested for safe operation.
CAUTION
Always unplug the product before cleaning or servicing.
Always unplug the product, set the brakes, and place blocks under the litter frame for support when working under
the product.
The recommended cleaners for this product's surfaces include the following:
Quaternary cleaners (active ingredient - ammonium chloride)
Phenolic cleaners (active ingredient - o-phenylphenol)
Chlorinated bleach solution (5.25% - less than 1 part bleach to 100 parts water)
Hand wash all surfaces of the product with warm water and mild detergent. Dry thoroughly.
Avoid oversaturation and make sure the product does not stay wet longer than the chemical manufacturer's guidelines
for proper disinfecting.
CAUTION
Always make sure that you wipe each product with clean water and thoroughly dry each product after cleaning.
Some cleaning products are corrosive in nature and may cause damage to the product if you use them improperly. If
you do not properly rinse and dry the product, a corrosive residue may be left on the surface of the product that
could cause premature corrosion of critical components. Failure to follow these cleaning instructions may void your
warranty.
Do not steam clean, pressure wash, ultrasonically clean, or immerse any part of the product in water. Exposure to
water may damage the internal electric parts. These methods of cleaning are not recommended and may void this
product's warranty.
®
Always clean Velcro
after each use. Saturate Velcro with disinfectant and allow the disinfectant to evaporate.
Appropriate disinfectant for nylon Velcro should be determined by hospital protocol.

Cleaning a support surface

To clean and disinfect a support surface, see the cleaning and disinfecting instructions in the support surface operations
manual.
www.stryker.com

Cleaning

2141-800-1 07 REV A
English
EN
1-117