Footboard LED indicators

EN
The LED indicators on the footboard control panel illuminate when there is a parameter change on the product.
LED
Bed unplugged
Bed charging
Brake not set
Battery low
Call maintenance
Locks enabled
1-38
Operation
Indicator
Illuminates amber when you unplug the product. Battery back-up functionality activates
when you unplug the product.
CAUTION
Plug the product into a wall outlet (regulated AC power source) when not in use to
maintain a sufficient battery charge and to maximize product performance while on
battery power.
Illuminates amber when you connect the product to a wall outlet and the batteries are
recharging. The batteries fully charge in approximately eight hours. When the batteries
are fully charged, the LED no longer illuminates.
If the batteries are not connected, or if the battery switch is in the OFF position (O), the
LED indicator does not illuminate.
Flashes amber when you release the brakes
Flashes amber when the battery charge is low
To recharge the InTouch batteries, plug the power cord into a properly grounded,
hospital-grade wall outlet.
CAUTION
Do not use the Zoom motorized drive when you hear a battery low alarm (Battery Low
LED on Footboard and audible beep). Stop using the Zoom motorized drive and
recharge the batteries immediately. If you ignore the battery low alarms, the batteries
may degrade quicker than normal and may decrease battery life.
Illuminates amber when the product requires maintenance or repairs. Contact the
appropriate maintenance personnel to restore proper functionality.
Illuminates amber on the footboard control panel when one lock or a total lockout is
set (see
Main menu: Lockouts on page
2141-800-1 07 REV A
1-86).