Carefully read and strictly follow the warnings and cautions listed on this page. Service only by qualified personnel.
WARNING
Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in
this manual.
Do not modify the product or any components of the product. Modifying the product can cause unpredictable
operation resulting in injury to patient or operator. Modifying the product also voids its warranty.
Always allow the product to reach room temperature before conducting any setup or testing functional operations to
prevent permanent product damage.
Always operate the product when all operators are clear of the mechanisms.
Always plug the product directly into a properly grounded, three-prong receptacle. You can only achieve grounding
reliability when you use a hospital-grade receptacle. This product is equipped with a hospital-grade plug for
protection against electric shock hazard.
Always properly handle the power cord to avoid the risk of entanglement, damage to the power cord, or potential
shock hazards. If the power cord is damaged, immediately remove the product from service and contact the
appropriate maintenance personnel.
Do not attach the power cord to any moving parts of InTouch.
Always unplug the power cord, turn the battery switch to the OFF (O) position, press the N/Brake Off button, and call
maintenance if unanticipated motion occurs.
Always unplug the product power cord from the wall when using oxygen administering equipment. Possible fire
hazard exists when this product is used with oxygen administering equipment other than nasal, mask type, or half
bed-length tent type.
The optional iBed
®
Wireless function is only intended to provide remote information of product status and parameter
conditions. It is not intended to replace patient monitoring protocol.
The line of sight between an iBed Locator and the head end of bed must be free of obstruction at all times. Any
interference could interrupt communication between the iBed Locator and the IR module.
®
You must use an iBed
Wireless compatible footboard with an iBed
®
iBed
Wireless functionality if you use an older version of the footboard.
You must verify all iBed
®
product's positions, and that iBed
the loss of information or the transmission of incorrect information.
You must install the iBed Locators more than 71" apart from one another in the same room (such as in a semi-
private room with more than one product). Failure to do may result in the product transmitting information to the
incorrect iBed Locator.
®
Do not use iBed
Wireless to replace the existing nurse call system. iBed
transmit product information. They are not intended to transmit nurse call information.
Always correctly associate or map the iBed Locator to the room or location to provide accurate location information.
Failure to properly map the iBed Locator to the room or location may result in the product transmitting incorrect
information.
If you move an iBed Locator after it has been installed and mapped, you must remap to the new room or location in
which it is moved to. You must also remap the iBed Locator if you change the room or location information after
initial installation.
Always wash your hands after handling a battery. Battery posts, terminals and related accessories contain lead and
lead compounds, chemicals known to the State of California to cause cancer and birth defects or other reproductive
harm. Properly dispose of batteries when required.
Power save mode activates after one hour on battery power with no motion release switch activation. Bed exit, scale,
and product motion stops operating when the product enters the power save mode.
Always keep feet clear from the area above the base cover or below the base cover when lowering the product or
when applying the brakes or releasing the brakes.
www.stryker.com

Summary of safety precautions

Wireless functionality after installation. Verify that the iBed locator communicates the
®
Wireless communicates with the wireless access point. Failure to do may result in
2141-800-1 07 REV A
®
Wireless compatible product. You will lose
®
Wireless products are only intended to
English
EN
1-13