WARNING
EN
Do not clean, service, or perform maintenance while the product is in use.
At a minimum, check all items listed during annual preventive maintenance for all Stryker Medical products. You may
need to perform preventive maintenance checks more frequently based on your level of product usage.
Remove product from service before performing preventive maintenance. Preventive maintenance should only be
performed by trained or certified personnel.
Notes
Clean and disinfect the exterior of the support surface before inspection, if applicable.
For Isolibrium preventive maintenance items, see the Isolibrium operations/maintenance manual.
Inspect the following items:
All welds and all fasteners are secure
Tubing or sheet metal for bends or breaks
Casters are free of debris
Casters are secure and swivel properly
Casters lock securely by depressing the brake pedal
Manual and electric brakes apply and release properly
Brake Not Set LED on the footboard and head end siderails when brakes are not engaged
Locking steer caster engages and disengages properly (model 2131)
Steer caster latches properly
Fowler operates properly
Litter up/down operates properly
Trend operates properly
IV pole is intact and operating properly (optional)
Support surface cover after each use
Support surface cover for rips or cracks
Headboard, footboard, and siderail panels for cracks or splits
All covers are not damaged and are not producing sharp edges
Night light operating properly
CPR release operating properly
All siderail motion functionality
Siderails move, latch, and stow properly
Siderail switches working properly (iBed Awareness option)
iBed Awareness light bars on footboard and siderails are working properly (iBed Awareness option)
All functions on head end siderails working properly (including LEDs)
All functions on footboard working properly (buttons, touch screen display, and LEDs)
Touch screen is accurately calibrated
Product is accurately calibrated
Scale and bed exit system operating properly
Drive wheel to make sure it is operating properly (Zoom motorized drive) (Model 2141) (option)
Motion release switches working properly (Zoom motorized drive) (Model 2141) (option)
Head end Zoom handle functionality operates properly (Zoom motorized drive) (Model 2141) (option)
Batteries for replacement (every two years) (use only QDF9188 for battery replacement)
Batteries for corrosion at the terminals, cracking, expanded or bulging at the sides, or can no longer maintain a
full charge
Nurse call functionality (optional)
Lubricate where required
Pendant for any physical damage
1-118