Power cord not frayed
Cables not worn or pinched
All electrical connections tight
All grounds secure to the frame
Ground impedance not more than 100 mΩ (milliohms)
Current leakage not more than 300 µA (microamps)
Ground chains are clean, intact, and have at least two links touching the floor
Enclosure is free from wear, tear, stresses and mechanical damage
No rust or corrosion of parts
Labels for legibility, proper adherence, and integrity
iBed Wireless Module and IR Module are intact and footboard icons are displaying (iBed Wireless option)
Product Serial Number:
Completed by: ______________________________________ Date:_____________
www.stryker.com
Preventive Maintenance
2141-800-1 07 REV A