Carefully read and strictly follow the warnings and cautions listed on this page. Service only by qualified personnel.
E E N N
WARNING
Improper usage of the product can cause injury to the patient or operator. Operate the product only as described in
this manual.
Do not change the product or any components of the product. Modifying the product can cause unpredictable
operation that results in injury to the patient or operator. Modifying the product also voids its warranty.
Risk of Shock. This product must only be connected to a supply mains with protective earth. Grounding reliability can
be achieved only when a hospital grade receptacle is used.
Explosion risk. This product is not suitable for use in the presence of flammable anesthetic mixture with air or with
oxygen or nitrous oxide other than nasal, mask type, or half bed-length tent type.
Do not move the support surface without assistance.
Do not operate the support surface without the Isolibrium top and bottom covers and linens.
Always position the support surface on the InTouch bed frame to that the support surface does not contact the
headboard.
Do not use when gaps between the support surface and the bed frame are present to prevent the risk of entrapment.
Only use the Isolibrium support surface with InTouch bed frames.
Do not attach the power cord to any moving parts of the InTouch bed frame. Improper routing of the power cord
could cause the risk of electric shock, pinching the power cord, or interfere with bed functions.
Always confirm proper functionality of the scale system after you install the support surface to prevent product
malfunction.
Do not use extension cords with Isolibrium. It is only intended to be powered by InTouch with the power cord
supplied.
Always use the power source provided by the InTouch foot end auxiliary power outlet with Isolibrium.
Always make sure that the operator has access to the CPR straps.
Do not stick needles into a support surface through the support surface cover. Holes may allow body fluids to enter
the inside (inner core) of the support surface and could cause cross-contamination, product damage, or product
malfunction.
Do not use fitted sheets with this support surface.
Always center the patient on the support surface. Align the patient's head toward the headboard before starting
functions. Check the patient frequently to make sure that you maintain the proper positioning.
Always make sure that the tubing and wiring that is connected to the patient is long enough, stable, and secure
during Lateral Rotation or Turn Assist.
Always raise all of the InTouch bed siderails before starting Turn Assist or Lateral Rotation functions.
Do not exceed the safe working load of the Isolibrium support surface. Excess weight could cause unpredictable
safety and performance of this system.
Always use extra caution when reading radiology images taken of a patient on a support surface because internal
components can cause artifacts and distort readings.
Always monitor the patient condition at regular intervals for patient safety.
Do not use the support surface as a transfer device to avoid the risk of patient injury.
Do not immerse the support surface or foot box in cleaning or disinfectant solutions.
Do not allow fluid to pool on the support surface. Fluids may cause degradation of components and may cause
unpredictable safety and performance of this product.
Always inspect the support surface covers (top and bottom) for tears, punctures, excessive wear, and misaligned
zippers as you clean the covers. If the cover is compromised, remove the support surface from service immediately
and replace the cover to prevent cross-contamination.
1-10

Summary of safety precautions

2971-109-001 REV B