WARNING
•
Always unplug the support surface power cord before cleaning or disinfecting to avoid the risk of shock.
•
Do not immerse the support surface or foot box in cleaning or disinfectant solutions.
•
Do not allow fluid to pool on the support surface. Fluids may cause degradation of components and may cause
unpredictable safety and performance of this product.
•
Always inspect the support surface covers (top and bottom) for tears, punctures, excessive wear, and misaligned
zippers as you clean the covers. If the cover is compromised, remove the support surface from service immediately
and replace the cover to prevent cross-contamination.
•
Always perform preventive maintenance based on the usage level of the product. An increase in usage which may
include more frequent cleaning and disinfection, may adversely affect the life of the support surface.
•
Always make sure that you wipe each product with clean water and thoroughly dry each product after cleaning or
disinfecting. Some cleaning and disinfecting agents are corrosive in nature and may cause damage to the product if
you use them improperly. If you do not properly rinse and dry the product, a corrosive residue may be left on the
surface of the product that could cause premature degradation of critical components. Failure to follow these
cleaning or disinfecting instructions may void your warranty.
CAUTION
•
Do not iron, dry-clean, or tumble dry the support surface covers.
•
Do not power wash the support surface.
•
Always completely dry the support surface covers before storing, adding linens, or placing a patient on the surface.
Prerequisite: Minimum of two operators are required to clean the bottom support surface cover.
Always follow hospital protocol for cleaning and disinfecting.
To clean the support surface covers between patient uses, follow these steps in order:
1.
If powered, unplug the InTouch from the wall outlet and turn the battery switch to off.
2.
Unplug the support surface power cord.
3.
Disconnect the support surface integration cable.
4.
Using a clean, soft, damp cloth, wipe the support surface covers with a mild soap and water solution to remove
foreign material.
5.
Wipe the support surface covers with a clean, dry cloth to remove any excess liquid or cleaning agent.
Note: To clean the bottom cover of the support surface, raise the head of bed and foot end to help fold the head
end and foot end of the support surface.
6.
Thoroughly rinse and dry the support surface covers.
www.stryker.com
Cleaning
2971-209-001 REV B
English