Cleaning
BED CLEANING
Hand wash all surfaces of the bed with warm water and mild detergent . Wipe the bed clean and DRY THOROUGHLY
to avoid build up of cleaning solution . Do not steam clean or hose off the bed . Do not immerse any part of the bed .
Some of the internal parts of the bed are electric and may be damaged by exposure to water .
Suggested cleaners for bed surfaces:
• Quaternary Cleaners (active ingredient - ammonium chloride) .
• Phenolic Cleaners (active ingredient - o-phenylphenol) .
• Chlorinated Bleach Solution (5 .25% - less than 1 part bleach to 100 parts water) .
Avoid over saturation and ensure the product does not stay wet longer than the chemical manufacturer's guidelines
for proper disinfecting .
CAUTION
SOME CLEANING PRODUCTS ARE CORROSIVE IN NATURE AND MAY CAUSE DAMAGE TO THE PRODUCT IF USED
IMPROPERLY . If the products described above are used to clean Stryker patient care equipment, measures must be
taken to ensure the beds are wiped with a damp cloth soaked in clean water and thoroughly dried following cleaning .
Failure to properly rinse and dry the beds will leave a corrosive residue on the surface of the bed, possibly causing
premature corrosion of critical components . Failure to follow the above directions when using these types of cleaners
may void this product's warranty .
MATTRESS CARE
Inspect the mattress after each use . Discontinue use if any cracks or rips are found in the mattress cover as these
openings may allow fluid to enter the mattress .