CPR MICRO SWITCH REPLACEMENT (OPTIONAL)

Tools Required:
• #1 Phillips Screwdriver
• #2 Phillips Screwdriver
N
H
Procedure:
1 .
Raise the bed to the highest position and apply the brakes .
2 .
Raise the head siderails .
3 .
Remove the plastic litter cover (if equipped) of the fowler or raise the fowler to the highest position .
4 .
Unplug the power cord from the wall receptacle .
FOWLER LEVER MICRO SWITCH
This switch informs on the state of the Fowler actuator (engaged or disengaged) when the CPR emergency release is
activated .
5 .
Using a #1 Phillips screwdriver, remove the two screws (A) holding the micro switch to the fowler lever . Keep the
mounting plate (B) . Remove the defective micro switch .
6 .
Remove the cables from the micro switch . Note the cable locations in order to properly reconnect them to the
new switch.
7 .
Reverse the above steps to install the new micro switch .
8 .
Test the CPR positioning before returning the bed to service .
FOWLER STRUCTURE MICRO SWITCH
This switch signals to the motor control board on the one hand that the resetting of the Fowler actuator may begin
because the Fowler is completely lowered, and on the other hand that the lowering to flat of the Gatch may also begin .
9 .
Remove the cable wires (C) from the micro switch . Note the location of the wires so they will be connected
correctly to the new micro switch.
10 . Press the switch clips to remove it from the support .
11 . Reverse the above steps to install the new micro switch .
12 . Test the CPR positioning before returning the bed to service .
www.stryker.com
Service Information
L
C
K
J
I
B
A
Figure 17
280C-009-002 REV B
M
A
A
E
F
D
Return To Table of Contents
L
B
B
G
1-59

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Gobed ii lit medsurgFl28c

Table des Matières