Pre Wash
Enzyme
Wash
Wash 1
Rinse 1
Dry Phase
Cleaning was validated using
Enzol® at 1 oz/gal.
Cleaning was validated using
Renu-Klenz™ at ¼ oz/gal.
Maintenance
Inspection and
Testing
Packaging for
Sterilization
Automated wash
2
Program the washer using the settings below.
Recirculation
Time
2 min.
2 min.
2 min.
2 min.
7 min.
Drying
3
Remove the tray from the washer after the
completion of the dry phase.
Dry the tray using a clean soft cloth. Use ltered
pressurized air (40 psi) to assist in drying.
Visually inspect the tray for cleanliness, paying
close attention to hard-to-reach areas.
No routine maintenance is required for the
sterilization tray.
Inspect the tray for dents and cracks. If a
problem is observed or suspected, the tray
should be returned for repair.
Inspect all components for cleanliness. If uid
or tissue buildup is present, repeat the above
cleaning and disinfection procedures.
Double-wrap the tray unless otherwise noted
below.
Water
Temperature
Cold tap water
Hot tap water
Set point 60°C
(140°F)
Heated 60°C
(140°F)
Drying Temperature
115°C (239°F)
Detergent
(type and
concentration)
Enzymatic
Detergent
Non-enzymatic
Detergent