MedComp DIGNITY Instructions D'utilisation page 13

Chambre d'injection implantable en plastique à injecteur automatique
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 27
Pro-Fuse Port/Catheter
Configuration
Low Profile
CT Ports
Standard
CT Ports
Non-Coring Needle
Gauge Size
Maximum Flow Rate
Note: CT injection pressure should be set at a maximum of 325 psi. Flow
rates less than 5 ml/s and/or lower viscosity contrast will generate lower
pressures in the port and catheter.
Represents flow capability of port and catheter assembly for power
a
injection of contrast media.
Internal port pressure during maximum indicated CT flow rate using
b
contrast media with 11.8 Centipoise (cp) viscosity.
Average static burst pressure and range is the burst pressure of the port
c
catheter assembly.
Dignity Mini
Thru
4.928 mm
Septum
Port
10.135 mm
Base
REFERENCES
1.
Jacobs, D.M. et. al., "Anatomical and Morphological Evaluation of
Pacemaker Lead Compression. PACE. 1993 Mar; 16(1):434-444.
2.
Magney, J.E. et. al., "Anatomical Mechanisms Explaining Damage
to Pacemaker Leads, Defibrillator Leads, and Failure of Central
Venous Catheters Adjacent to the Sternoclavicular Joint". PACE.
1993 Mar; 16(1):445-457.
3.
Hinke, D.H.; Zandt-Stastny, D.A.; Goodman, L.R.; et. al. "Pinch-off
syndrome: A complication of implantable subclavian venous access
devices". Radiology 177: 353-356, 1990.
4.
Ingle, Rebecca; Nace, Corinne. "Venous Access Devices: Catheter
Pinch-off and Fracture." 1993, Bard Access Systems, Inc.
5.
Camp-Sorrell, Dawn. "Access Device Guidelines.: 2nd Ed. Oncology
Nursing Society, 2004.
FURTHER READING
See Medcomp
CT Guide for more details.
Contact a Medcomp
information about any of these products.
Medcomp® WARRANTS THAT THIS PRODUCT WAS MANUFACTURED ACCORDING TO
APPLICABLE STANDARDS AND SPECIFICATIONS. PATIENT CONDITION, CLINICAL
TREATMENT, AND PRODUCT MAINTENANCE MAY EFFECT THE PERFORMANCE OF
THIS PRODUCT. USE OF THIS PRODUCT SHOULD BE IN ACCORDANCE WITH THE
INSTRUCTIONS PROVIDED AND AS DIRECTED BY THE PRESCRIBING PHYSICIAN.
Because of continuing product improvement, prices, specifications, and model
availability are subject to change without notice. Medcomp
to modify its products or contents in accordance with all relevant regulatory
requirements.
Triniflex
is a registered trademark of Medical Components, Inc.
®
Does not contain DEHP.
Average Port
Reservoir
Pressure
5F
116 psi
6.6F
69 psi
8F
83 psi
8F
87 psi
9.6F
44 psi
Indicated CT Flow Rates
19 Ga.
5ml/s
a
Needle Penetration Depth
Dignity
Low Profile
4.928 mm
10.135 mm
Implantable Infusion Port Patient Guide and/or
®
Sales Representative for more
®
WARRANTY
Average
Static Burst
Pressure
b
249 psi
237 psi
228 psi
197 psi
143 psi
20 Ga.
5ml/s
Dignity
Mid-Size
6.477 mm
11.354 mm
-11-
Static Burst
Pressure
Range
c
232-262 psi
232-245 psi
218-242 psi
177-225 psi
141-144 psi
22 Ga.
2ml/s
Pro-Fuse
Pro-Fuse
Low Profile
Standard
4.928 mm
6.096 mm
10.135 mm
11.633 mm