Preventive Maintenance - Stryker 2971 Isolibrium Manuel D'utilisation

Surface de soutien. destinée à une utilisation avec le châssis de lit intouch fl27 (2131, 2141, 2151, 2152, version logicielle 4.0 ou supérieure).
Masquer les pouces Voir aussi pour 2971 Isolibrium:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

WARNING
Do not service or perform maintenance while the product is in use.
At a minimum, check all items listed during annual preventive maintenance for all Stryker Medical products. You may
need to perform preventive maintenance checks more frequently based on your level of product usage. Service only by
qualified personnel.
Remove product from service before you perform preventive maintenance inspection.
Note: Clean and disinfect the exterior of the support surface before inspection, if applicable.
Inspect the following items:
Zipper, snaps, and covers (top, bottom, and fire barrier) are free of tears, cuts, holes, or any other damage
Note: If excessive wear is observed on the cover or the fire barrier, it is strongly recommended to replace the
cover.
Support surface cover labels are legible, have proper adherence, and free of damage.
Handles are free of rips or cracks
Power cord, integration cable, and plug are free of damage
**Perform a functional test of the Isolibrium functions
**Perform system diagnostics and confirm there are no errors
*Internal components for signs of staining from fluid ingress or contamination by fully unzipping the cover
Pods are not leaking or cracking
Hose connections to the manifold and sensor tubes (both ends) are seated fully
Foam is free from large tears or large gouges
Turn bladders are free of excessive wear and function properly
Low Air Loss fans rotate freely and are free of debris and dust
Left and Right CPR releases function properly
Both foot box cooling fans rotate freely and are free of debris and dust
Pump spring mount has visible clearance from the foot box bottom
Listen for abnormal pump vibration and noise level
Clean the foot box intake metal screen
Clean the mesh on the bottom cover
Clean the fan filter screen on the foot box
Standoffs (feet of the foot box) are present
Cable tie is present on the power cord
Current leakage not more than 200 µA
Ground impedance not more than 100 mΩ
Note: The ground lug can be accessed through the slit in the cover for the CPR, patient left.
www.stryker.com

Preventive maintenance

(Figure 1-9 on page
Figure 1-9: Ground lug (patient left side)
2971-209-001 REV B
1-19)
English
EN
1-19