Bottom connectors
figure 3: Bottom connectors
leD Indicators
DC In
Fault
figure 4: leD Indicators
UsInG the vessIx Generator systeM
1. Assemble the Vessix Cart System per the included instructions. Then mount the generator and power supply onto the
roll cart per the System Assembly Instructions included with the packaged Vessix Generator System. Connect the
power supply and patient extension cable to the generator. Plug the power cord into an AC outlet.
2. Turn the Vessix Generator on by pressing the Power Button. The system will power up in Standby Mode (power on,
no RF output) displaying the splash screens (Figure 5).
3. The screen will advance to the "Waiting for Connection Screen" (Figure 6), at which time the patient extension cable
and catheter should be connected to the bottom of the generator.
connector
UsB connector
Data Port
Dc In connector
Vessix Power Supply
5V
6.7A
Dc In Indicator
fault Indicator
figure 5: splash screens
figure 6: "Waiting for connection" screen
5
Supplies RF power output to Vessix
Reduce™ Catheter. Catheter may also
optionally connect directly to this port
through the patient extension cable.
Boston Scientific service use only.
Connects mains power from DC power
supply to generator. Only use the
IEC 60601-1 approved power supply
provided with the Vessix™ Generator.
Lights to indicate whether the generator
is connected.
Lights to indicate that a fault has
occurred. Photograph screen or write
down the Alert Code and two sub-
codes, turn system off and contact
Boston Scientific.