MR Conditional System Description
Supervision
Note: The patient should be in a psychological condition and mental state in which the patient is able to
provide immediate feedback of any problems during the examination. Maintain visual and audio
monitoring of the patient throughout the MRI examination. Verify that the patient is feeling normal
and is responsive during and between each individual scan sequence of the MRI examination.
Discontinue the MRI immediately if the patient becomes unresponsive to questions or experiences
any adverse effects listed in the Safety Information Section of this manual under Potential
Interactions with MRI Environment.
Post-MRI Examination Review
Verify that the patient has not experienced any adverse effects as a result of the MRI. The potential adverse
effects are listed in the Safety Information Section of this manual under Potential Interactions with MRI
Environment. Contact Boston Scientific if the patient has experienced any adverse effects.
ImageReady™ MRI Guidelines for Boston Scientific Deep Brain Stimulation Systems
91098813-02 Rev C
8 of 277