Prep-C: Inspect the Device
IFU 5686 AE, 2024-04
Perform the following inspection to prevent premature Collagen exposure:
1. Black Sleeve is locked in position.
2. Collagen Patch is not exposed.
3. Key is not engaged in the Lock and the Key is located at the proximal end of the device.
Check the Disc function:
1. Firmly hold the Silver Handle.
2. Pull back the Black Actuator until it locks in place.
3. The Green Segment will be visible.
Check the Disc shape:
1. Circular appearance.
2. Symmetrical appearance.
3. Intact membrane.
If there are any device defects, do not use the device. Report and return to the
manufacturer.
Collapse the Disc by pressing the Black Actuator tip. Device is ready for use.
Page 14 of 24