CAUTION – Device is restricted to sale to or on the order of a physician.
Description
The VASCADE MVP® Venous Vascular Closure System (VVCS) is
intended to seal the femoral venous access site(s) after an
endovascular procedure. The system consists of a sterile disposable
Vascular Closure Catheter, which houses a resorbable Collagen Patch
and a sterile Clip (see Figure 1).
The system is designed to deliver a resorbable Collagen Patch
extra-vascularly at the vessel puncture site to aid in achieving
haemostasis. The patch expands as a result of rehydration in the
presence of blood in the tissue tract to provide an extravascular
seal. A radiopaque proximal marker band on the Catheter provides
means to verify placement of the patch in the tissue tract adjacent
to the femoral vessel site before the release of the patch. A second
distal marker band locates the distal tip of the VASCADE Disc. In
addition, the Disc is echogenic. The following version of the
VASCADE MVP VVCS is available:
For use in 6F to 12F ID (max 15F OD) 12 cm
•
sheaths
Indications For Use
The VASCADE MVP Venous Vascular Closure System (VVCS) Model
800-612C is indicated for the percutaneous closure of femoral
venous access sites while reducing time to ambulation, total post-
procedure time, time to haemostasis and time to discharge eligibility
compared to manual compression, and enabling same-day discharge
in patients who have undergone catheter-based procedures using
6–12F inner diameter (15F maximum outer diameter) procedural
sheaths, with single or multiple access sites in one or both limbs.
Contraindications
The device should not be used in patients with a known allergy to bovine derivatives.
Note:
Clinicians are responsible for informing patients in advance of the procedure that the collagen in the implant is derived
•
from animal tissue.
•
The materials used in the device do not contain or consist of carcinogenic, mutagenic or reprotoxic (CMR) substances
or endocrine disruptors.
1
Overall length of the sheath (including the hub) needs to be less than 15 cm.
IFU 5686 AE, 2024-04
INSTRUCTIONS FOR USE
en
VASCADE MVP® Venous Vascular Closure System (VVCS)
6–12F (Venous)
1
introducer
GREEN
SEGMENT
(VISIBLE
WHEN
DISC IS
DEPLOYED)
INSERTION
POINT
PROXIMAL
MARKER BAND
DISTAL
MARKER BAND
Figure 1. VASCADE MVP VVCS
MD
BLACK ACTUATOR
SILVER HANDLE
ORANGE˜BLUE KEY
LOCK
BLACK SLEEVE
WHITE MARKER STRIPE
COLLAGEN PATCH
(HOUSED UNDER
BLACK SLEEVE)
DEPLOYED DISC
Page 1 of 24