FOOTBOARD CONTROL MEMBRANE REPLACEMENT

Tools Required:
• Small Regular Screwdriver
• #2 Phillips Screwdriver
B
C
Procedure:
1 .
Raise the bed to the highest position and apply the brakes .
2 .
Unplug the power cord from the wall receptacle .
3 .
Using a small regular screwdriver, lift and remove the five round self-adhesive screw-covers (A) located on the foot
board cover and the self-adhesive membrane (B) of the foot board control panel to expose the screws holding the
cover to the foot board . Use caution when inserting the screwdriver under the membrane to avoid scratching the
plastic cover .
Note
Do not reuse the self-sticking parts removed since their self-adhesive coating considerably looses its efficiency once
they are removed . Replace them with new parts .
4 .
Properly ground yourself (see the "Static Protection Procedure",
5 .
Using a #2 Phillips screwdriver, remove the seven screws (C) holding the cover to the foot board and remove the
cover after having removed the cables connected to it . Note the cable positions in order to replace them properly .
6 .
Remove the defective membrane .
7 .
Reverse the above steps to install the new membrane .
Note
Use caution to avoid adhering the cable to the back of the membrane .
8 .
Test all the functions of the foot end control panel before returning the bed to service .