K-MINC-2000 Instructions for Use
Guidance and manufacturer's declaration – Electromagnetic emissions
The MINC+ Benchtop Incubator is intended for use in the electromagnetic environment specified below. The customer or the end user of the MINC+
Benchtop Incubator should ensure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 32
Harmonic emissions
Class A, Class D
IEC 61000-3-2
Conducted emissions
Group 1
CISPR 11
Class B limits
Conducted emissions
Class B limits
CISPR 32
Voltage fluctuations/
flicker emissions
Complies
IEC 61000-3-3
Proximity fields from RF wireless communication equipment
Test frequency
Band
(MHz)
(MHz)
385
360-390
450
430-470
710
745
704-787
780
810
870
800-960
930
1,720
1,845
1,700-1,990
1,970
2,450
2,400-2,570
5,240
5,500
5,100-5,800
5,785
Note: If necessary to achieve the IMMUNITY test level, the distance between the transiting antenna and the ME equipment or ME system can be
reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
1. For some services, only the uplink frequencies are included.
2. The carrier shall be modulated using a 50% duty cycle square wave signal.
3. As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does not represent actual modulation, it would be
worst case.
WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such
use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
WARNING: Use of cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions
or decreased electromagnetic immunity of this equipment and result in improper operation.
WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than
30 cm (12 inches) to any part of the MINC+ device, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment
will result.
Electromagnetic environment guidance
The MINC+ Benchtop Incubator uses RF energy only for its internal function. Therefore, its RF emissions are
very low and unlikely to interfere with nearby electronic equipment.
The MINC+ Benchtop Incubator is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low-voltage power supply network that supplies buildings used
for domestic purposes.
Service
Modulation
1
2
Pulse
TETRA 400
modulation
2
18 Hz
FM
3
GMRS 460
± 5 kHz deviation
FRS 460
1 kHz sine
Pulse
LTE band 13,17
modulation
2
217 Hz
GSM 800/900
TETRA 800
Pulse
iDEN 820
modulation
2
CDMA 850
18 Hz
LTE band 5
GSM 1800
CDMA 1900
Pulse
DECT
modulation
2
LTE band 1, 3,
217 Hz
4, 25; UMTS
Bluetooth
WLAN
Pulse
802.11 b/g/n RFID
modulation
2
2450
217 Hz
LTE band 7
Pulse
WLAN 802.11
modulation
2
a/n
17 Hz
Maximum
Distance
power
(m)
(W)
1.8
0.3
2
0.3
0.2
0.3
2
0.3
2
0.3
2
0.3
0.2
0.3
1-32
English
Immunity
Minimum
test level
separation distance
(V/m)
(m)
27
0.3
28
0.3
9
0.3
28
0.3
28
0.3
28
0.3
9
0.3
1