K-MINC-2000 Instructions for Use
5. A message will appear: 'Database restore has been completed
successfully'.
6. Press OK to return to the Database tab. The USB drive can be
removed from the port.
5.4.21 Upgrade DishTrace MINC+ software
1
The DishTrace MINC+ software may occasionally be required to be
updated. DishTrace MINC+ is published by William A. Cook Australia Pty.
Ltd. and MINC+ will automatically confirm the validity of the software
during installation. If the publisher is different or cannot be verified
by the device, contact your service agent. An option to downgrade
the DishTrace MINC+ software to the previous version is available,
however should only be used under instruction from Cook Medical or an
authorised service agent.
To update the DishTrace MINC+ software:
1. Insert the USB drive containing the software upgrade files into the
USB port on the front of the MINC+.
2. Tap the Settings screen and then tap the Database tab.
3. Next to the Upgrade Software, tap Upgrade. A message will
appear asking for conformation to upgrade the software.
4. Tap Yes to continue with the upgrade. The MINC+ will reboot once
the software update is complete.
5.4.22 Restore factory settings
WARNING: Selecting Restore Factory Settings will delete all logged
data and restore default settings.
The Restore Factory Settings function deletes all dish and patient
records. Performing this function before requesting a service operation
is recommended. This function also resets the DishTrace MINC+ function
settings to the default.
1. Tap the Settings screen and then tap the Database tab.
2. Tap Reset, which is next to Restore Factory Settings. A message
will appear: 'All data including user settings, database will be
deleted. Do you want to proceed anyway?'
3. Tap Yes to continue. The device settings will be restored, and the
center console will restart automatically.
4. After the device restarts, the date and time and possibly network
settings will need to be reconfigured. See Section 4.12 for details.
5.4.23 Information – Device settings
This screen shows information about the incubator. There are no user
actions on this screen.
• Incubator Name – Incubator name.
• Serial No. – Incubator serial number.
• BIOS Version – Operating system BIOS version for center console.
• OS Version – Operating system version for the center console.
• DishTrace MINC+ Version – DishTrace MINC+ software version
number.
• FW Version – The firmware version for the MINC+ core control system.
• Manufacture Date – The date on which the device was manufactured.
• Last Serviced Date – The date on which the device was last serviced.
5.4.24 DishTrace MINC+ screen saver
The screensaver is a Cook Medical image that appears on the center
console touchscreen every 6 hours when the touchscreen is not used.
It appears on the center console touchscreen for only a few seconds to
reduce image burn-in. The screensaver cannot be modified or disabled.
The two outer incubator chamber screens also display a black screen as
a screensaver for a few seconds, every 6 hours, if the chamber controls
have not been used. The screensavers will not affect the operational
aspects of the MINC+.
5.5 DishTrace PC software
The DishTrace PC software is a comprehensive software tool that shows
the status of multiple MINC+ incubators connected to a local network.
It shows a list and the status of all the active MINC+ devices currently
connected as well as any previously connected incubators that may be
currently offline. Offline status may indicate that the device is turned off
or has been sent away for servicing.
The DishTrace PC software must be installed in accordance with Section
4.13 to establish a secure connection between the PC with DishTrace PC
installed and each MINC+ on the network. Once the connection process
is completed, each device will be automatically detected when turned on
and added to the list of active incubators. If a MINC+ is later turned off, it
will still appear on the list but will be shown as inactive on the network.
For each incubator, DishTrace PC displays the status of the following:
• Incubator chambers
• Humidification flasks
• Gas
• Dish records
• MINC+ administration information
5.5.1 Using the DishTrace PC software
To run DishTrace PC, do one of the following:
• Click on the DishTrace PC shortcut on the desktop.
• Click on the Windows operating system start button, scroll down to the
Cook Medical folder, and click on the DishTrace PC file.
The main screen will appear, and the Incubators screen will be shown by
default.
5.5.2 Incubators screen
The Incubators screen is shown by default when DishTrace PC is
launched. Any active incubator on the network will be automatically
connected and shown.
5.5.3 Adding incubators
If an incubator fails to appear on the list after several minutes have
elapsed, it can be manually added:
1. On the top left-hand side, in the MINC+ address bar (as shown in the
screenshot above), enter the IP address of the incubator you want to
add. (To find the IP address on the MINC+ incubator, go to Settings
> Network.)
2. Press the Add button. After the incubator is found, a message will say
that the incubator has been added successfully. Click OK.
5.5.4 Incubator display
The total number of online and offline incubators is shown on the
top right-hand side. The incubator list can be shown as a graphical
representation or a list view, either of which can be selected by clicking
any of the icons next to the VIEW label.
To view further details of an incubator, double-click on any incubator
entry. The following screen will appear:
1-21