K-MINC-2000 Instructions for Use
8. Technical data
Classification according to IEC 60601-1
Type of protection against electric shock
Degree of protection against harmful ingress of solids and water
General specifications
Power supply
Frequency
Overvoltage category
Pollution degree
Mains supply voltage fluctuations
Maximum power consumption
Alarm contact rating
Environmental operating conditions
Storage and transport directions
Manufactured and tested to the following standards
Dimensions
Weight (device only)
Gas supply type
Gas supply pressure
Gas flow rate, nominal
Gas flow rate, accuracy
Chamber temperature capability
Incubation chamber temperature accuracy
Media warming chamber
Computer requirements for the DishTrace software
Operating system
Processor and RAM
Free hard disc space
Processor
Display
I/O ports
Specifications
Maximum number of connected MINC+ devices
Maximum logging time period
Log file format
Report format
USB drive format for backup and reports
Graph update interval
Maximum displayed time period on screen
Temperature resolution
Class I equipment
IP31
100 - 240 VAC
50 - 60 Hz
II
2
+/-10%
200 VA
30V
2.0A
Indoor use only: + 18 °C to + 32 °C at temperature set point of 37.0 °C
+ 20 °C to + 28 °C for all other temperature settings
10% to 85% RH
700 hPa to 1,060 hPa (Altitude up to 3000m)
+ 5 °C to + 40 °C
10% to 85% RH
IEC 60601-1: 2005 + AMD1:2012
IEC 60601-1-2: 2014
IEC 61010-1: 2010 + AMD1:2016, IEC 61010-2: 010: 2019
UL 61010-1 (3rd Ed.): Am.1
CSA C22.2 No. 61010-1 3ED + Amd 1
CSA C22.2 No. 61010-2-010:15
Door closed: 556 mm wide x 197 mm high x 449 mm deep.
Door opened: 556 mm wide x 528 mm high x 449 mm deep.
23.4 kg (51.6 lb)
Blend of 6% CO
, 5% O
, 89% N
(Cook Culture System at sea-level) or high
2
2
2
purity 6% CO
in air (recommended tolerances ± 0.2%).
2
150 kPa ± 15 kPa (21.8 psi ± 2.2 psi) (1,500 mbar ± 150 mbar)
25-29 mL/min per chamber (both chambers in normal flow status).
Gas-purge flow rate at 300 mL/min per chamber for 3 minutes (both chamber
in purge flow status).
4 mL/min per chamber (normal flow).
± 30 mL/min per chamber (purge).
35.0 °C to 40.0 °C in 0.1 °C increments in an ambient temperature range of
+ 20 °C to + 28 °C. At set point of 37.0 °C, the ambient temperature range is
extended to + 18 °C to + 32 °C.
± 0.2 °C at calibration point.
Temperature set point range 35.0 °C to 40.0 °C. Accuracy +/- 1 °C.
Windows 8.1, Windows 10.
Minimum configuration as specified by the host OS.
10 GB minimum.
Pentium 4, 1.6 GHz or better.
1024 x 768, 256 color or better.
Ethernet port for connecting to networked MINC+ devices.
USB port for connecting up to ten K-MINC-1000 incubators via hub.
USB port for installation of software.
50 (including ten K-MINC-1000)
Limited only by free disc space.
Comma-separated values (CSV)
PDF
FAT32
25 seconds
5 days
0.1 °C
1-30
English
1