K-MINC-2000 Instructions for Use
Buttons and
Description
features
Media warming
7
MWC temperature status indicator
chamber lights
Keypad unlock (times out; will relock after 5
8
1
seconds of no keypad input)
USB port to download patient and dish
9
USB port
information and for service maintenance updates
10
Center console touchscreen
11
Enable/disable gas flow
Enable/disable monitoring of the water level in the
12
humidification flask
13
Increase/decrease chamber temperature set point
14
Enable/disable the incubation chamber heater
Display icons
Description
Incubation chamber temperature set point
Open incubation chamber lid
Incubation chamber heater active
Incubation chamber heater inactive
Gas flow normal (flowing at a steady rate)
Gas purge (actuates after the lid closes to
reestablish environmental conditions)
Gas flow disabled
Water level monitoring disabled
Water level empty
Water level full
4.4 Rear of the device
2
3
1
4
Figure 4: Rear incubator controls
Description
1 Type B USB socket (for service only)
2 Center console reset (for DishTrace MINC+)
External alarm contacts (either normally open [NO] or
3
normally closed [NC])
4 IEC power inlet socket
5 Gas inlet port
6 Ethernet connection socket
6
5
4.5 Electromagnetic compatibility
The MINC+ Benchtop Incubator complies with the electromagnetic
compatibility (EMC) limits for medical devices as specified by IEC 60601-
1-2: 2014. These limits provide reasonable protection against harmful
interference in a typical medical environment.
Medical electrical equipment requires special precautions regarding
EMC and must be installed and operated according to these instructions.
It is possible that high levels of radiated or conducted radio-frequency
electromagnetic interference (EMI) from portable and mobile RF
communications equipment or other strong or nearby radio-frequency
sources could result in performance disruption. Evidence of disruption
may include erratic readings, equipment ceasing to operate, or other
incorrect functioning. If any of these disruptions occur, cease using the
incubator and contact your Cook Medical service agent.
For guidance and the manufacturer's declaration on electromagnetic
emissions and immunity of the incubator, refer to Section 8.
4.6 Incubator placement
The MINC+ should be placed on a secure level surface away from heaters,
coolers, air-conditioning outlets, mists, splashes, and exposure to direct
sunlight. Position the device to ensure that it can be easily disconnected
from the power supply. Once installed and placed into service, the
MINC+ is not intended to be moved from one place to another. The lab
environment should be clean and the device must be kept away from
areas with dust.
Allow a reasonable amount of space around each MINC+ to ensure
sufficient air circulation. At least 1 cm space is recommended around
each device. Allow at least 60 cm of headroom, measured from the
bottom of the incubator, to allow the lid to open.
WARNING: The MINC+ should not be used adjacent to or stacked with
other equipment or be stacked during storage. If adjacent or stacked use is
necessary, please ensure that appropriate shelving racks are used to allow the
device to function normally.
WARNING: EXPLOSION HAZARD. Do not use the MINC+ in the
presence of flammable gases.
WARNING: ELECTRIC SHOCK HAZARD. Do not immerse the MINC+.
WARNING: ELECTRIC SHOCK HAZARD. Determine if the available
voltage corresponds with the MINC+. Connecting the device to the wrong
voltage will cause the incubator to malfunction or may permanently damage
the device. The power cord must be equipped with a safety plug. Use the
enclosed power cord for the connection between the power point and the
device socket.
WITHIN CANADA & USA – Use only a listed detachable power-supply cable,
type SJT, minimum 18 AWG x 30, with 3 conductors, and one end configured
for NEMA 5-15 and the other end for IEC 320/CEE22.
To avoid the risk of electric shock, this equipment must be connected to a
supply main with protective earth.
1-7
English