• Use only hospital-approved disinfectants to clean the adapter. The following solutions are acceptable for cleaning or disinfecting
the adapter: water, Betadine™, hydrogen peroxide, and mild detergent. After cleaning, wipe the adapter with a sponge or soft cloth
and water.
• Immediately remove any contamination from the adapter by using a sponge or soft cloth and water.
• If contamination cannot be removed with a sponge or soft cloth and water, the adapter may be immersed in water for up to
10 minutes. Do not immerse the adapter in any other liquids.
11 Service
Warning: The adapter is not serviceable. Do not disassemble the adapter because the performance of the adapter may be adversely
affected. If the adapter has been tampered with, do not use it.
12 Specifications for the Affinity CP centrifugal blood pump with Cortiva bioactive surface
Part number:
Priming volume:
Inlet/Outlet ID:
Maximum operating pressure:
Maximum flow rate:
Maximum outlet pressure:
Maximum pump speed:
Compatible systems:
Store the product at room temperature.
12.1 Transport range
Temperature:
Humidity:
13 Specifications for the Affinity CP adapter
Part number:
Compatible pump:
Compatible systems:
Maximum adapter speed:
IP rating:
13.1 Gross dimensions
Weight:
Height:
Diameter:
13.2 Storage and transport environment
Temperature:
Humidity:
Atmospheric pressure:
English
CBAP40
40 mL (approximate)
9.5 mm (3/8 in)
760 mmHg (101 kPa)
10 L/min
700 mmHg (93.3 kPa)
• 4000 rpm (with Medtronic external motor drive 560A)
• 3600 rpm (with Affinity CP adapter on Stöckert and Sorin centrifugal pump systems or
Sarns and Terumo centrifugal systems)
Medtronic speed controller systems (with external motor drive model 560A, emergency hand-
crank model HC150A)
-30°C to 40°C (-22°F to 105°F)
20% to 85% noncondensing
AP40AST
Affinity CP centrifugal blood pump with Cortiva bioactive surface (CBAP40)
• Stöckert and Sorin centrifugal pump systems (with drive unit models 60-01-04 and
60-01-00 and emergency drive unit models 60-01-35 and 60-01-50)
• Sarns and Terumo centrifugal systems (drive unit model 164267 and manual drive unit
model 164268)
3600 rpm
IP47
315 g (0.69 lb)
28 mm (1.1 in)
94 mm (3.7 in)
−30°C to 70°C (−22°F to 158°F)
15% to 93% noncondensing
70 kPa to 106.3 kPa