INSTRUCTIONS FOR USE
All caregiver controls, along with their indicators, operate correctly; this includes
controls on the Allen Advance Table and on the pendant.
The motors operate correctly: during articulation, there is no friction, abnormal noises,
or an audible indication of overload.
The battery is in good condition: with the unit unplugged, the pendant controls operate
all Allen Advance Table articulations; the battery is charged.
The brakes operate correctly.
The casters are in good condition; they do not have cuts, are not worn, and have a
good amount of tread.
The leakage current is less than 100 μA.
The accessories are in good condition.
3.7
Parts Replacement
WARNING:
Only facility-authorized maintenance persons should do these parts replacement
procedures. Otherwise, patient fall, injury, or equipment damage could occur.
WARNING:
Always wear the applicable personal protective equipment when you service
the unit. Failure to do so could cause injury.
Document Number: 80028198
Version: F
Controls and Motors
Battery
Brakes and Casters
Electrical Checks
Accessories
Page 79
English
Issue Date: 03 FEB 2023
Ref Blank Template: 80025117 Ver. F