INSTRUCTIONS FOR USE
8.
Do as follows to unlock the support tops for the flip:
a. Lift the end of the flip handle (A) until it locks in
position.
b. Push and hold the red safety lever (B) forward as you
pull the flip handle (A) down to the released position.
NOTE: The manual flip indicator on the display turns orange to
show that the support tops are no longer locked in position
and they can be flipped. The pendant indicator on the display
turns off to show that you can not use the pendant controls
while the Allen Advance Table is in flip mode.
9.
After the patient has been flipped, and the tops are level
(the tilt level indicator on the display is green), push the flip
handle up to the locked position. You do not need to
press the red safety lever.
NOTE: The flip indicator on the display turns off to show that the support tops are locked
in position. The pendant indicator on the display turns green to show that you can use the
pendant controls.
10. Make sure the support tops are locked in position. If the tops did not lock, repeat
Step 9.
11. Remove the straps from around the support tops.