INSTRUCTIONS FOR USE
• The air pump is not connected to the air manifold at the head end of the
lateral top.
• The lateral top electrical connector is not connected to the auxiliary power
port.
2.
Install the prone support bracket.
a. Make sure that the hook feature of the side support bracket is over the pin on the
main support bracket then turn the main support bracket until the cross-pin hole
in the side support bracket aligns with the cross-pin hole in the main support
bracket.
WARNING:
Do not use the Lateral-to-Prone Assembly without the cross pin installed correctly.
To do so could cause patient fall, injury, or equipment damage.
b. Insert the cross pin into the prone support bracket, through the lateral support
bracket, and through the other end of the prone support bracket. Make sure that
the drop-lock end of the cross pin is down.
c. Make sure the drop lock on the end of the cross pin
moves to the side to hold the H-bracket in position.
d. Repeat Step a through Step c at the other end of the
table.
3.
Install the rail top with the prone support set as follows:
a. Make sure that the rail top has the supine access tops installed, and not the sling.
If necessary, remove the sling and install two supine access tops. Refer to "Install
the Prone Support Set (A-71300) on the Rail Top" on page 46.
b. At each end of the rail top, install a cross pin into the
prone bracket, through the rail top connector
assembly, and the other side of the prone bracket.
Make sure that the cross pins that secure the rail top
have the drop-lock end down.