Télécharger Imprimer la page

Hillrom Allen A-71100-CA Mode D'emploi page 14

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 453
INSTRUCTIONS FOR USE
1.6
EMC considerations:
WARNING:
If the Allen Advance Table is to be used near high-frequency surgical equipment,
cardiac defibrillators, and cardiac defibrillator-monitors, refer to the product
manufacturer's instructions to make sure the equipment will operate as intended.
This device meets all requirements for electromagnetic compatibility per
IEC 60601-1-2. Other products that are used near the Allen Advance Table should
comply with this standard also. Otherwise, electromagnetic interference
between the products could cause the products to operate incorrectly. If
problems do occur, contact the product manufacturer(s).
WARNING:
The A-71100-XX/A-71101-XX and A-71705-XX should not be used adjacent to
other electrical equipment. If adjacent use is necessary, observe the A-71100-XX/
A-71101-XX or A-71705-XX and the other electrical equipment to make sure they
operate as intended.
WARNING:
This device meets all requirements for electromagnetic compatibility per
IEC 60601-1-2. It is unlikely that the user will encounter problems with this device
because of inadequate electromagnetic immunity. However, electromagnetic
immunity is always relative, and standards are based on anticipated
environments of use. If the user observes unusual device behavior, particularly if
such behavior is intermittent and associated with nearby use of radio or TV
transmitters, cell phones, or electro-surgical equipment, this could be an
indication of electromagnetic interference. If such behavior occurs, the user
should try to move the interfering equipment further from this device.
Make sure the A-71100-XX/A-71101-XX and A-71705-XX operate correctly when
they are used near other electronic devices. Portable and mobile radio
frequency (RF) communications equipment can affect electrical equipment.
Medical equipment needs special precautions in regard to electromagnetic
compatibility (EMC) and needs to be installed and put into service according to
the EMC information supplied in the tables that follow.
Document Number: 80028198
Version: F
Page 14
English
Issue Date: 03 FEB 2023
Ref Blank Template: 80025117 Ver. F