3.5.1
The Allen Advance Table will not go into Pendant Control Mode from the Manual
Flip Mode
Make sure of these:
• The brakes are locked.
• The columns are at the same height.
• The Allen Advance Table and the installed support top are level "Level the Allen
Advance Table and the Support Top" on page 54.
• The Allen Advance Table is free of objects that could interfere with the adjustment.
• The flip handle is in the up and locked position.
If all of the conditions above are met, and the Allen Advance Table still will not go into the
manual flip mode, reset the power: press the On/Off switch to off and then to on.
3.5.2
The Pendant Control and Manual Flip Indicators are Flashing
Make sure that the flip handle is fully in the up and locked position or down and released
position. If the indicators continue to flash, reset the power: press the On/Off switch to off
and then to on.
3.5.3
The Brakes Locked and Unlocked Indicators are Flashing
Press and hold either the Brakes Lock or Unlock pendant control until the applicable
Brakes locked/unlocked indicator on the display stays on; the indicators are no longer
flashing.
3.5.4
The Allen Advance Table "Beeps" when a Pendant Control is Pressed
A "beep" from the Allen Advance Table lets you know that the battery charge level is low
or depleted. Plug the Allen Advance Table into an applicable power outlet to charge the
batteries. To keep the batteries charged, plug the Allen Advance Table into an outlet
nightly and when you store the Allen Advance Table.
3.5.5
Service Calls
When you call Allen Medical about your unit, be prepared to give the serial number from
the product identification label. You will find the label on the inside of the head-end base
near the on/off switch.