3.2.11
Change the Height of the Advance Pinless H-Bracket
WARNING:
When you lift up on the Advance Pinless H-Bracket handles, the caregivers or
assistants must support the weight of the patient and the patient support top. Use
a minimum of three people at each Advance Pinless H-Bracket to support the
weight of the patient and the patient support, and operate the Advance Pinless
H-Bracket. Injury and equipment damage can occur.
WARNING:
Observe lines closely during articulations. Always use good line management
techniques. Failure to do so could cause patient fall, injury, or equipment
damage.
WARNING:
Make sure that you keep hands, fingers, tubes, and lines clear of the center of the
bracket when you adjust the height of the patient support surface. Failure to do
so could cause patient fall, injury, or equipment damage.
1.
Use a minimum of three people at each Advance Pinless H-Bracket to raise the
Advance Pinless H-Bracket handles to adjust the height of the patient support.
a. One person should stand on the patient's left side to support the weight of the
patient and patient support top. One person should stand on the patient's right
side to support the weight of the patient and patient support top. The third person
should stand at the head or foot end as applicable and operate the lift triggers.
2.
Lift up gently on the patient support so that the weight of the patient and the patient
support is held by the caregivers or assistants, not by the Advance Pinless H-Bracket.
WARNING:
Make sure that there are no objects between the patient support surface and the
floor before you lower the patient support surface or adjust the table. Failure to
do so could cause patient fall, injury, or equipment damage.
3.
Press and hold the lift trigger to release the mechanism. Each lift trigger will operate
the other lift trigger.