INSTRUCTIONS FOR USE
12. Remove the lateral top as follows:
WARNING:
At least two people are necessary to safely remove the lateral top. The lateral top
can fall when the cross pins are removed. Always have a second person present
to make sure that the lateral top does not fall. Failure to do so can cause patient
injury and equipment damage.
a. Disconnect the lateral top electrical connector.
b. Have a second person hold the lateral top so that it does not fall when the cross
pin is removed.
c. Make sure that the drop lock on the end of the cross pin aligns with the cross pin,
and remove the cross pin at each end of the lateral top. Remove the cross pin at
the foot end of the lateral top first.
d. Remove the lateral top.
13. Install the arm boards on the rail top.
14. Remove the Drawsheet from the patient.
15. Put the patient in the prone position as directed by the surgeon.
3.2.16
Flip the Support Top 180°
WARNING:
Before you adjust the Allen Advance Table or use the flip or tilt features, make
sure that the brakes are locked, the patient's weight is equally distributed on the
support top, the patient is fully prepared for the adjustment; the Allen Advance
Table is free of objects that could interfere with the adjustment, and all loose or
unsupported objects have been secured. Otherwise, patient fall, injury, or
equipment damage could occur.
1.
Make sure that—
• The brakes are locked (the locked indicator on the
Allen Advance Table display is green).
• The patient is secured to the support top and is fully
prepared for the adjustment.
Document Number: 80028198