INSTRUCTIONS FOR USE
5.
Pull the assembly out from the column, and make sure
the assembly is in the locked-out position. The Cervical
Traction Release button will return to its original position.
WARNING:
The maximum diameter of the traction rope should
be 0.25" (6.35 mm). The use of a traction rope that is
greater than 0.25" (6.35 mm) in diameter could
cause binding inside the cervical traction assembly.
Patient injury could occur as a result of inaccurate
traction.
6.
Put the traction rope through the carriage, and through
the hole in the center of the bowtie. The rope will go
through the traction assembly and out the left side of
the column, near the release.
7.
Put the traction rope over and around the pulley.
8.
Press the carriage adjustment releases and put the
carriage in the desired position. See "Change the
Cervical Traction Angle" on page 70.
WARNING:
The maximum weight load for the traction assembly
is 50 lb (22.6 kg). To exceed this weight could cause
patient fall, injury, or equipment damage.
WARNING:
Make sure the traction weight is in a safe position, free from possible unintended
movement. Failure to do so could cause patient fall, injury, or equipment
damage.
Document Number: 80028198
Version: F
Page 69
English
Issue Date: 03 FEB 2023
Ref Blank Template: 80025117 Ver. F