Télécharger Imprimer la page

Hillrom Allen A-71100-CA Mode D'emploi page 16

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 453
INSTRUCTIONS FOR USE
Guidance and Manufacturer's Declaration — Electromagnetic Immunity
The A-71100-XX/A-71101-XX and A-71705-XX are intended for use in the
electromagnetic environment specified below. The customer or the user of the
A-71100-XX/A-71101-XX or A-71705-XX should make sure it is used in such an
environment.
IEC 60601
Immunity Test
Test Level
Electrostatic
± 6 kV Contact
Discharge (ESD)
± 8 kV Air
IEC 61000-4-2
Electrical Fast
± 2 kV for Power
Transient/Burst
Supply Lines
± 1 kV for Input/
Output Lines
IEC 61000-4-4
Surge
± 1 kV Line(s) to
Line(s)
± 2 kV Line(s) to
IEC 61000-4-5
Earth
Voltage Dips,
< 5% U
Short
(> 95% dip in U
Interruptions,
for 0.5 cycles
and Voltage
40% U
Variations On
(60% dip in U
Power Supply
for 5 cycles
Input Lines
70% U
(30% dip in U
IEC 61000-4-11
for 25 cycles
< 5% U
(> 95% dip in U
for 5 seconds
Document Number: 80028198
Version: F
Compliance
Level
± 6 kV Contact
± 8 kV Air
± 2 kV for Power
Supply Lines
± 1 kV for Input/
Output Lines
± 1 kV Line(s) to
Line(s)
± 2 kV Line(s) to
Earth
< 5% U
T
T
)
(> 95% dip in U
T
for 0.5 cycles
40% U
T
T
)
(60% dip in U
T
for 5 cycles
70% U
T
T
)
(30% dip in U
T
for 25 cycles
< 5% U
T
T
)
(> 95% dip in U
T
for 5 seconds
Page 16
English
Electromagnetic Environment—
Guidance
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
)
of a typical commercial or
T
hospital environment. If the user
of the A-71100-XX/A-71101-XX
or A-71705-XX requires
)
T
continued operation during
power mains interruption, it is
recommended that the
)
A-71100-XX/A-71101-XX or
T
A-71705-XX be powered from
an uninterruptedly power
supply or a battery.
)
T
Ref Blank Template: 80025117 Ver. F
Issue Date: 03 FEB 2023

Publicité

loading