INSTRUCTIONS FOR USE
4.5
Cleaning and Disinfection Instruction:
We recommend that you clean and disinfect the Allen Advance Table and the support
tops between patient use. Refer to your facility's cleaning and disinfection policies, and
follow the guidelines below.
WARNING:
• Do not use bleach or products that contain bleach to clean the device. Injury
or equipment damage can occur.
• After each use, clean the device with alcohol-based wipes.
• Do not put the device into water. Equipment damage can occur.
• Use a cloth and a quaternary ammonium disinfecting/cleaning solution to
clean and disinfect the device.
• Read and follow the manufacturer's recommendation for low-level
disinfection.
• Read and follow the cleaning product's instructions. Use caution in areas
where liquid can get into the mechanism.
• Wipe the device with a clean, dry cloth.
• Make sure that the device is dry before you store it or use it again.
• The potential for electrical shock exists with electrical equipment. Failure to
follow facility protocols may cause death or serious injury.
• Failure to unplug the unit from its power source could cause patient fall, injury,
or equipment damage.
• Always wear the applicable personal protective equipment when you clean
the unit. Otherwise, injury could occur.
• Do not put the unit in an equipment wash room. Patient fall, injury, or
equipment damage could occur.
• Do not expose the unit or its components to excessive moisture. Patient fall,
injury, or equipment damage could occur.
• Do not use harsh cleaners, solvents, or detergents. Equipment damage could
occur.
NOTE: Do not use iodophors; they will stain the product.
• If you use a quaternary-type disinfectant, make sure to rinse it off the product
immediately. Otherwise, the disinfectant could cause corrosion on painted
and plated components.