387158--G_GA_Lumax540_mul.book Page 24 Friday, February 1, 2013 5:38 PM
ICD leads
Connections
Pacing/sensing
• VR-T , VR-T DX: 1 x IS-1 bipolar
• DR-T: 2 x IS-1 bipolar
• HF-T: 3 x IS-1 bipolar
ICD lead configuration
Connection Lead connector
RV
DF-1 connector of the ventricular shock coil
SVC
DF-1 connector of the vena cava electrode
S A
IS-1 connector of the atrial sense electrode, bipolar
P/S A
IS-1 connector of the atrial pace/sense electrode, bipolar
P/S V
IS-1 connector of the ventricular pace/sense electrode, bipolar
CS
IS-1 connector of the ventricular pace/sense coronary sinus electrode,
bipolar or unipolar
Tolerances
The following tolerances apply to the programmable parameters:
Parameter
Pacing pulse amplitudes
Pacing pulse width
Atrium sensitivity
DR-T, HF-T
Atrium sensitivity
VR-T DX
Sensitivity of right and left ventricle 40%
Pacing intervals
Escape interval
Detection intervals
Refractory periods
ATP coupling interval
All manuals and user guides at all-guides.com
Shock
• All device types: 2 x DF-1
Tolerance
+20% to -25%
10%
For values  0,4 mV: 0.2 mV to 0.52 mV
For values  0.4 mV: +30% to -50%
For values  0,4 mV:
–0.1 mV to +0.3 mV; for values  0.4 mV:
+30% to –60%
20 ms
20 ms
20 ms
20 ms
20 ms
Parameter
Shock coupling interval
Shock energy/charging voltage
Included Equipment Lumax
Standard
In the sterile container there is:
• Device of the Lumax product family
• Blind plug DF, mounted in the header
• Blind plug IS-1, mounted in the header (only for HF devices)
• Screwdriver
In the storage package there is:
• Patient's ICD manual
• Sterile cover for the PGH programming head of the programmer
• Implantation protocol
• Serial number label
• Patient ID card
• Follow-up report
• Warranty card
• Technical manual
Orders
The devices are available with the following order numbers (some device types are not
available in all countries):
Device type
540 VR-T
540 VR-T DX
540 DR-T
540 HF-T
Accessories
All accessories must correspond to the requirements of the EC Directive 90/385/EEC.
• BIOTRONIK leads
• BIOTRONIK programming and monitoring devices
• Permanent magnet M-50
• For Home Monitoring: BIOTRONIK transmitters
24
Tolerance
20 ms
Electrical characteristics: see peak voltages
Order number
360348
368852
360346
360347