387158--G_GA_Lumax540_mul.book Page 22 Friday, February 1, 2013 5:38 PM
Resistance to interference
With unipolar sensing, the requirements of EN 45502-2-2:2008 are met for interference
voltages  0.6 mV (peak to peak).
EMC requirements relating to section 27.5 are met as long as atrial sensitivity is set to
1.0 mV (corresponds to factory settings) or higher ( 1.0 mV). Appropriate measures
must be taken to assure interference-free therapy if more sensitive values are chosen.
Common mode rejection
Indication of rate in Hz, of common mode rejection in dB:
Rate
Atrium
Atrium
VR-T DX
VR-T, DR-T, HF-T
CMRR (Common Mode Rejection Ratio)
16.6
67
65
50
80
80
60
81
75
ATP amplitude
Set parameters:
Burst
Amplitude
Pulse width
R-S1 interval
Number S1
Load resistance
Results:
ATP amplitude Value with tolerance Measured minimum Measured maximum
RV
7.5 V 1,5 V
Mean value
LV
7.5 V 1.5 V
Mean value
BiV
7.5 V 1.5 V
Mean value
All manuals and user guides at all-guides.com
V right
V left
64
64
67
64
66
63
Value
7.5 V
1.5 ms
300 ms
5
500 
7.42 V
7.50 V
5.18 V
5.18 V
7.50 V
7.50 V
5.18 V
5.18 V
7.42 V
7.50 V
5.16 V
5.18 V
Sensitivity setting
Indication of minimum and maximum of automatic sensitivity setting; measurement of
the actual values with positive and negative polarity:
Note: The programmed atrial sensitivity is intensified by a factor of 5 for 
Lumax 540 VR-T DX.
Sensitivity
Test signal wave
shape
A: positive [mV]
2
sin
15 ms
Standard triangle
A: negative [mV]
2
sin
15 ms
Standard triangle
VR-T DX: A: positive
2
sin
15 ms
[mV]
Standard triangle
VR-T DX:
2
sin
15 ms
A: negative [mV]
Standard triangle
RV: positive [mV]
2
sin
40 ms
Standard triangle
RV: negative [mV]
2
sin
40 ms
Standard triangle
LV: positive [mV]
2
sin
40 ms
Standard triangle
LV: negative [mV]
2
sin
40 ms
Standard triangle
22
Value
Tolerance
Measured
value
0.2
0.2 ... 0.5
0.26
0.41
0.2
0.2 ... 0.5
0.28
0.43
0.2
0.02 ... 0.10
0.08
(0.1 - 0.5)
0.08
0.2
0.02 ... 0.10
0.10
(0.1 - 0.5)
0.12
0.5
0.3 ... 0.7
0,56
0.58
0.5
0.3 ... 0.7
0.45
0.58
0.5
0.3 ... 0.7
0.64
0.65
0.5
0.3 ... 0.7
0.54
0.68

Publicité