Forced termination; with SMART, SMART redetection
Parameter
Range
Forced termination
OFF 1 ... (1) ... 15 min
Redetection counter
Parameter
Redetection counter VT1
Redetection counter VT2
Therapy Parameters
Ventricle
Parameter
Range
Energy of 1st shock and 2nd
OFF 1 ... (1) ... 16 ... (2) ... 40 J
shock VT1, VT2
Energy of 1st shock and 2nd
1 ... (1) ... 16 ... (2) ... 40 J
shock VF
Shock sequences 3rd-nth shock
OFF
in VT1
1*40 J; 2*40 J; 3*40 J; 4*40 J;
5*40 J; 6*40 J
Shock sequences 3rd-nth shock
OFF
in VF
4*40 J; 5*40 J; 6*40 J
Number of shocks
0 ... (1) ... 8
(VT1/VT2)
Number of shocks (VF)
2; 6 ... (1) ... 8
Confirmation in VT1, VT2, VF
OFF; ON
Shock form in VT1, VT2, VF
Biphasic; biphasic2
Polarity in VT1, VT2, VF
Normal; inverse; alternating
Shock path (valid for all shocks
RV SVC + housing
including the painless shock
RV housing
impedance)
RV SVC
Shock path DX
en • English
All manuals and user guides at all-guides.com
Standard
1 min
Range
Standard
10 ... (2) ... 30
20
10 ... (2) ... 30
14
Standard
40 J
40 J
6*40 J
6*40 J
8
8
ON
Biphasic
Normal
RV SVC +
housing
RV housing
ATP parameters
Parameter
ATP type
ATP attempts
Ventricular pacing
(HF devices only)
Number S1
Add. S1
R-S1 interval
S1 decrement
Scan decrement
S1-S2 interval
Minimum ATP interval
ATP timeout
ATP optimization
ATP pulse amplitude
ATP pulse width
ATP One Shot parameters; ATP in VF
Parameter
ATP type
ATP attempts
Ventricular pacing
(HF devices only)
Number S1
19
Range
Standard
Burst; Ramp; Burst + PES
Burst
OFF
3
1 ... (1) ... 10
RV; LV; BiV
RV
1 ... (1) ... 10
5
OFF; ON
ON
200 ... (10) ... 500 ms (absolute);
80%
70 ... (5) ... 95% (adaptive)
5 ... (5) ... 40 ms
10 ms
OFF
OFF
5 ... (5) ... 40 ms
200 ... (10) ... 500 ms (absolut);
70%
70 ... (5) ... 95% (adaptive)
200 ... (5) ... 300 ms
200 ms
OFF
OFF
00:15 ... (00:15) ... 05:00 mm:ss
OFF; ON
OFF
7.5 V
7.5 V
1.5 ms
1.5 ms
Range
Standard
OFF;
OFF
Burst; Ramp; Burst + PES
1
1
RV; LV; BiV
RV
1 ... (1) ... 10
8