Device type
Pacing modes
VR-T
VVI, VVIR,
OFF
VR-T DX
VDD, VDDR,
VDI, VDIR,
VVI, VVIR
OFF
Timing Parameters Lumax DR-T
Basic rate day/night
Parameter
Range of values
Basic rate
30 ... (5) ... 100 ... (10) ... 160 ppm
Night rate
OFF
30 ... (5) ... 100 ppm
Night begins
00:00 ... (1 min) ... 23:59 hh:mm
Night ends
00:00 ... (1 min) ... 23:59 hh:mm
Rate hystereses
Parameter
Range of values
Rate hysteresis
OFF
-5 ... (-5) ... -90 ppm
Repetitive hysteresis
OFF
1 ... (1) ... 15
Scan hysteresis
OFF
1 ... (1) ... 15
AV delay
Parameter
Range of values
AV delay
Low; medium; high; fixed; individual Low
AV delay 1
15 (fixed); 40 ... (5) ... 350 ms
at rate 1
30 ... (10) ... 150 ppm
en • English
All manuals and user guides at all-guides.com
Standard
Parameter
VVI
AV delay 2
at rate 2
VVI
Sense compensation
Safety window
AV hystereses
Parameter
AV hysteresis mode
Standard
60 ppm
AV hysteresis
OFF
Positive repetitive AV hysteresis
22:00:00 hh:m
m
Negative repetitive AV hysteresis OFF
06:00 hh:mm
AV scan hysteresis
Standard
IRSplus
OFF
Parameter
IRSplus
OFF
AV hysteresis at IRSplus
OFF
Repetitive AV hysteresis at IRSplus
AV scan hysteresis at IRSplus
Standard
Post-ventricular atrial refractory period
Parameter
–
60 ppm
PVARP
13
Range of values
40 ... (5) ... 350 ms
40 ... (10) ... 160 ppm
OFF
-5 ... (5) ... -60 ms
100 ms
Range of values
OFF
Positive; Negative; IRSplus (only for
DR devices)
10 ... (10) ... 150 ms
OFF
1 ... (1) ... 10
1 ... (1) ... 15 ... (5) ... 100 ... (10) ... 180
OFF
1 ... (1) ... 10 ms
Range of values
OFF; ON
400 ms
OFF
1 ... (1) ... 10
OFF
1 ... (1) ... 10
Range of values
AUTO
175 ... (25) ... 600 ms
Standard
–
130 ppm
-30 ms
100 ms
Standard
OFF
50 ms
OFF
OFF
OFF
Standard
OFF
400 ms
5
5
Standard
250 ms