387158--G_GA_Lumax540_mul.book Page 20 Friday, February 1, 2013 5:38 PM
Parameter
Range
R-S1 interval
200 ... (10) ... 350 ms (absolute);
70 ... (5) ... 95% (adaptive)
S1 decrement
5 ... (5) ... 40 ms
S1-S2 interval
200 ... (10) ... 350 ms (absolute);
70 ... (5) ... 95% (adaptive)
Stability
12%
ATP pulse amplitude
7.5 V
ATP pulse width
1.5 ms
Progressive course of therapy
Parameter
Range
Progressive course of therapy
OFF; ON
Post-shock pacing
Parameter
Range
Pacing mode
• DDI when permanent DDD(R),
(automatically programmed)
• VDI when permanent VDD(R),
• VVI when permanent VVI(R);
• OFF
Basic rate
30 ... (5) ... 100 ... (10) ... 160 ppm
Rate hysteresis
OFF
-5 ... (-5) ... -65 ppm
AV delay
50 ... (10) ... 100 ... (10) ... 160 ms
Ventricular pacing
RV; BiV
LV T-wave protection
OFF; ON
Triggering
OFF
RVs; RVs + PVC.
Post-shock duration
OFF
00:10 ... (00:10) ... 00:50; 01:00 ...
(01:00) ... 10:00 mm:ss
All manuals and user guides at all-guides.com
Standard
85%
10 ms
70%
12%
7.5 V
1.5 ms
Standard
ON
Standard
DDI(R), AAI(R)
VDI(R)
60 ppm
OFF
140 ms
RV
ON
OFF
00:10 mm:ss
Diagnostics Functions
Home Monitoring, Holter Memory and Statistics
Home Monitoring
Parameter
Home Monitoring
Time of transmission
IEGM for therapy episodes
IEGM for monitoring episodes
Periodic IEGM
Sustained atrial episodes (DR, HF) OFF
Thoracic impedance
Holter episodes
Parameter
Range
AT/AF
OFF; ON
SVT
OFF; ON
Statistics
Parameter
Range
AT/AF rate
100 ... (10) ... 250 bpm
Start resting period
0:00 ... (10) ... 23:50 hh:mm
Resting period duration
0:00 ... (10) ... 23:50 hh:mm
Automatic impedance
VR, DX, DR: ON; OFF
measurement
HF: ON; OFF; LV OFF
Thoracic impedance
ON; OFF
20
Range
Standard
OFF; ON
OFF
00:00 ... (1 min) ... 23:59 hh:mm
01:00 hh:mm
OFF; ON
OFF
OFF; ON
OFF
OFF
OFF
2; 3; 4; 6 months
12 h
0.5; 6; 12; 18 h
ON; OFF
OFF
Standard
ON
ON
Standard
200 bpm
02:00 hh:mm
04:00 hh:mm
ON
OFF