Precision™ Spinal Cord Stimulator System Clinician Manual
Note: If you accidentally locate the patch in the wrong place, or if the Charging Belt moves out
of alignment, the Charger will start beeping again. Use a new adhesive patch or readjust
the belt to place the Charger back into position.
5. When the Charger emits a series of double beeps, the Stimulator is fully charged. Turn off
the Charger, remove the Charging Belt or adhesive patch, and return the Charger to the
Base Station. Do not confuse the end of charge signal (a series of double beeps) with the
continuous beeps that indicate that the Charger is searching for the Stimulator.
Note: • Depending on your program parameters, you may expect daily recharging times from
as low as 10 minutes up to four hours, or weekly recharging times from as low as one
hour up to four hours. Charging times may be longer than 4 hours when not properly
aligned.
• The end of a charge signal is a distinct double beep, and the alignment indicator is a
steady continuous signal.
• Remote will not be able to communicate with the IPG when charging.
Charger Battery Service Life
The charger battery has a typical service life of 500 charging cycles.

Patient Remote Control

Caution
• Keep the Remote Control dry. It should not be exposed to moisture.
• Keep the Remote Control away from pets, pests and children to avoid damage to the device.
Basic Operation
The Remote Control communicates with the Stimulator through a radio frequency (RF)
telemetry link from a distance of up to 60 cm (2 feet).
The Remote Control goes into idle (or sleep) mode when not in use. It can be reactivated by
pressing any button. To unlock the Remote Control, hold down
appears.
Clinician Manual
91173656-02
54 of 385
P
, until "Release P To Unlock"