3. Select either the Program tab or the Area tab, as appropriate.
4. Adjust the amplitude limit (Min or Max mA) using the up and down arrows. This will be the
minimum or maximum amplitude that can be used with the current Program or Area.
5. Select OK.
To specify the Ramp Up for the currently active Program or a specified Area:
1. Select
next to the Program Name.
2. Select Options.
3. Select the Program tab or the Area tab.
4. Check the Ramp Up check box to enable a ramped start.
5. Choose a duration for Ramp Up using the up and down arrows.
6. Select OK.

Specifying Remote Control Locks

The lockout selection provides or denies patient's access to the following features in the Remote
Control:
• Pulse Width control
• Rate Control
• Main Menu
• Area Menu
• Program and Schedules Menu
To lock out one of these features in the Remote Control:
1. Select
next to the Program Name.
2. Select Options.
3. Select the Stimulator tab.
4. Check the features you would like to deny access to in the Remote Control. The patient will
only have access to the features that are not checked.
Programming the Patient
Bionic Navigator™ 3D 4.0 Programming Manual
92395591-02 45 of 1595