Precision™ Spinal Cord Stimulator System Clinician Manual
Remove needle stylet
6. OPTIONAL. Under fluoroscopic guidance, insert the lead blank through the LOR adaptor
and into the epidural space. Advance the lead blank to verify entry into the epidural space,
then withdraw the blank.
7. The LOR adaptor must be removed prior to inserting the lead. Pull the LOR adaptor straight
out from the slotted needle without twisting or bending.
Recommended: Hold the slotted needle in place while removing the LOR adaptor.
Note: An anchor must be loaded over the distal end of the Infinion CX Lead prior to
inserting the lead into the Entrada Needle.
8. While holding the lead stylet handle, place the steering cap over the proximal end of the
stylet handle with moderate force until it is held in place. The lead stylet should be fully
inserted into the lead. Slowly insert the Infinion CX Lead into the slotted needle, directing
the distal lead tip into the needle lumen.
Clinician Manual
91173656-02
22 of 385
Verify entry into the epidural space.