Specifications And Technical Data; Materials - Boston Scientific Precision SC-1110 Mode D'emploi

Système de stimulation médullaire
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Precision™ Spinal Cord Stimulator System Clinician Manual

Specifications and Technical Data

Parameter
Areas (Channels)
Amplitude
Rate
Width
Cycle
Ramp ON
Contacts
a. The Precision System includes programmable coverage areas with each individ ual electrode contact limited to 12.7 mA.
A programming interlock is enforced to limit the coverage area output current to 20 mA or less. For example, a maximum
current output of 12.7 mA on a first electrode would limit the total summed current output on remaining electrodes to 7.3
mA within one coverage area.
b. Only one Area is available if the rate is 130 pps.

Materials

Case
Header
Strain Relief
Size/Volume
Clinician Manual
91173656-02
68 of 385
Range
4
0 – 20 mA
a
2 – 1200 pps
b
0 – 1000 µsec
0 – 90 min, OFF
1 – 10 secs
1 – 16; +, -, OFF
Titanium
Epoxy
Silicone
55 mm x 45 mm x 11 mm / 20.7 cm
Default
0 mA
40 pps
210 µsec
OFF
3 secs
1 – 16: OFF
(including header)
3