Lead, Lead Extension, And Splitter Handling And Storage - Boston Scientific Precision SC-1110 Mode D'emploi

Système de stimulation médullaire
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

• Do not incinerate an IPG. Improper disposal of the device could result in an explosion.
Devices should be explanted in the case of cremation, and returned to Boston Scientific
Neuromodulation. An explant kit is available.
Store the IPG between 0 °C and 45 °C (32 °F and 113 °F). Devices should always be kept in
temperature regulated areas within the acceptable temperature range. IPG damage can occur
at temperatures outside of this range.

Lead, Lead Extension, and Splitter Handling and Storage

• Avoid damaging the lead with sharp instruments or excessive force during surgery.
• Do not sharply bend or kink the lead, extension, or splitter.
• Do not tie suture(s) directly to the lead, extension, or splitter body; use the provided suture
sleeves.
• For the Percutaneous Leads - avoid forcing the lead into the epidural space by carefully
clearing a path using the lead blank.
• Avoid pulling an implanted lead taut; provide a stress relief loop at the insertion site to
minimize tension on the lead.
• Avoid handling the lead with sharp instruments; use only rubber-tipped forceps.
• Take care when using sharp instruments such as hemostats or scalpels to prevent
damaging the lead.
• Wipe off any body fluids from the lead connector end before connecting it to any other
component. Fluid contamination of these connections could compromise the integrity of the
stimulation circuit.
• Wipe off any body fluids from the lead stylet before inserting or reinserting it into the lead.
Store components between 0 °C and 45 °C (32 °F and 113 °F) in an area where they are not
exposed to liquids or excessive moisture. Temperatures outside of the stated range can cause
damage.
Sterilization, Handling, and Storage
Clinician Manual
91173656-02
15 of 385

Publicité

Table des Matières
loading

Table des Matières