Precision™ Spinal Cord Stimulator System Clinician Manual

Connecting to the IPG

For Percutaneous Lead - Dual Lead Connection

• Superior (upper or left) leads, extensions, or 2x4 splitters connect to IPG port "1-L".
• Inferior (lower or right) leads, extensions, or 2x4 splitters connect to IPG port "2-R".
For Infinion™ 16 Lead Connection
• Laser-marked tail of Splitter 2x8 connects to IPG port "1-L".
• Unmarked tail of Splitter 2x8 connects to IPG port "2-R"
Upper
Lower
2R
1L

For Infinion CX Lead Connection

• Proximal tail with one marker band connects to IPG port "1-L".
• Proximal tail with two marker bands connects to IPG port "2-R".

For Percutaneous Lead - Single Lead Connection

• Connect a single lead, extension, or 2x4 splitter to IPG port "1-L".
• Insert port plug into unused port.
Clinician Manual
91173656-02
46 of 385